September 2008 Training Manual: KI 7600 Series Power Meters Level 1, V 2. 0 20080825 1
Course Contents 1. 2. 3. 4. 5. General Features Models Detector Types Instrument Care Prepare instrument 6. 7. 8. 9. Modes of operation Display Modes Built in visible laser (KI 7601) Memory Operation 2
1. / General Features Ø Low skill operation Ø Autotest capability Ø Autotest compatible with Agilent N series instruments Ø Calibrated at multiple wavelengths Ø 1% factory calibration accuracy Ø Supplied with calibration certificate Ø Industry standard connectors – Including SFF Ø Ø Ø Ø Ø 3 High contrast LCD display with backlight Battery life – 190 -360 hours DC power socket Memory USB Computer interface Test tone detection Built in VLS variant (KI 7601) 3 year re-calibration interval 3 year warranty
2. / Models There are two model styles: Ø KI 7600: the most popular. Ø KI 7601: fitted with a Visible source on left hand port There are two Computer interface styles: Ø Early instruments: RS 232 port Ø Current instruments: USB Type A port RS 232 & USB models have minor operational differences 4
3. / Detector Types Choose detector to suit your application. In. Ga. As: Telco & LAN Ø Enhanced In. Ga. As with good response @ 850 nm Ø Power Levels of up to +5 d. Bm Ø Most accurate detector type at Telco wavelengths H series: High power – CATV, DWDM Ø Enhanced H 3 with good response @ 850 nm Ø Power level configurations of up to +27 d. Bm Ø Filtered In. Ga. As Other detector types available for specific wavelength and power requirements. 5
a. / Typical Detector Spectral responses 6
4. / Instrument Care Keep the instrument in its carry case during storage and transport Ø Use only high quality 1. 2 -1. 5 volt batteries. (Do not use lithium batteries) Ø For prolonged storage remove batteries. Ø The instrument is resistant to normal dust and moisture, however it is not waterproof. Ø If moisture gets into the instrument, remove batteries & dry it carefully before reuse Ø Where possible, keep instrument away from strong sunlight. Ø Clean the instrument case using Iso-Propyl-Alcohol (IPA) or other non solvent cleaning agents. Ø Ensure instrument input power rating is not exceeded DO NOT use Acetone or other active solvents as damage may result. 7
5. / Prepare Instrument a) Hand carry strap b) Control buttons c) Power supply – batteries & external d) Turn On / Off e) Fit / Remove adaptors f) Test cord selection 8
a. / Hand carry strap If required, fit provided hand carry strap: Ø Provides enhanced on-the-job instrument care 9
b. / Control buttons Computer Port Hidden Keypad DC Power Turn On/Off RS 232: Back Light USB: Memory store Detector Optional VLS Function buttons λ toggle Battery compartment 10
c. / Power supply - batteries To Install Batteries. Ø Hold instrument in both hands with battery compartment uppermost and with thumbs resting on battery compartment latch. Ø Press latch down and push away from case. Ø Insert batteries. Ø Insert ‘AA’ cells using the supplied AA-C battery converters. Battery life. Alkaline ‘C’ batteries : 360 hours meter, 190 hrs VFL Alkaline ‘AA’ batteries : approx 75 hours meter Low Battery Display. Indicator shows when approximately 10 hours left. 11
Power Supply - external All Instruments: Ø External power supply disconnects the batteries Ø Rechargeable batteries must be removed for charging RS 232: Ø Via the DC power socket USB: DC Ø Via the DC socket or via the USB cord B US Plug Pack Requirements: Ø 2. 5 mm DC power plug Ø 6 -12 V DC @ 300 m. A maximum Ø +Ve pin 12
d. / Turn On / Off 10 minutes auto Off or Permanent operation Push the oval [POWER] button to turn ON or OFF Low battery indicator To defeat auto time out. Hold down [POWER] for 3 seconds during turn ON 13
e. / Adaptor - fitting Industry standard SC adaptors Ø We recommend ceramic/zirconia sleeves Slot to outside 14
Adaptor - removal Current models: Ø Ø Locate quick release button on rear of instrument at base of connector housing Push and hold button in Pull out existing adaptor Fit new adaptor OR Ø Remove as per ‘early models’ Early models: Ø Move adaptor interface to mid position Ø Pull out existing adaptor Ø Fit new adaptor 15
f. / Test cord selection Power meters accept PC and APC connectors. KI 7601 has built in visible laser source (VLS) Ø Must specify visible source to be PC or APC when ordering Instrument ports are colour coded: VLS connector: Ø PC housing: BLUE Ø APC housing: GREEN Specify PC or APC VLS 16
6. / Modes of Operation 4 main modes of operation a) Autotest: Ø automatically toggles between all λs Ø Preferred mode for loss testing as testing time is greatly reduced. Ø Minimises error as meter always displays correct λ. b) Manual: Ø Single λ operation Ø Preferred mode for level monitoring. c) Modulated: Ø Displays incoming modulation frequency d) KITS software: Ø Under software control 17
a. / Autotest Simplest mode for loss testing When receiving light from a compliant source operating in Autotest mode, the meter will auto toggle between λs Ø Power meter receives data which contains wavelength, source serial number and nominal source output power. Ø If power meter not calibrated at an incoming wavelength it will ignore it but remain in sync with other wavelengths. Ø If incoming power level too low at a particular wavelength it will ignore it but remain in sync with other wavelengths. 18
b. / Manual Mode If not in power meter mode Ø Press [Power Meter] To select wavelength Ø Toggle [-/+]. Note: Ø Wavelength toggle is not circular Ø Most meters have the common wavelengths grouped together for speed of access when used in manual mode. E. g. 850, 1300, 1310, 1550, 1625 nm are together 19 [-/+]
c. / Test Tone When receiving test tone or low level modulation Ø Power meter displays modulated frequency and beeps Ø Built in feature cannot be disabled 20
d. / KITS™ Direct Interface to - KITS™ Testing & Reporting software Instrument under computer control Ø Familiar Excel™ User Interface Ø Inbuilt multi language support Ø Memory extract to Excel spreadsheet or CSV file Ø One Click Real Time Data Capture Ø Standards based & user definable analysis Ø Data Logging Ø User Customisable reports Ø Fee for service user customisation service 21
7. / Display Modes a) Absolute d. Bm / Relative d. Br mode b) Setting the Reference c) Log d. Bm / Linear W mode 22
a. / Absolute / Relative Mode Absolute Mode: Measure actual power level at a particular location – d. Bm Relative Mode: Measure power level ‘relative’ to a particular location - d. Br -6. 99 d. Bm -47. 71 d. Bm 0 d. Br -40. 72 d. Br 23
LCD Display λ Main display features - Absolute Mode: Ø Calibrated wavelength – nm Ø Received power - d. Bm Ø Nominal ‘Autotest’ source power d. Bm Ø Only if Autotest source Received Power Nominal Source Power λ Main display features - Relative Mode: Ø Calibrated wavelength - nm Ø Relative received power - d. Br Ø Reference power - d. Bm Reference Power 24 Relative Power
Toggle Absolute / Relative Mode If not in power meter mode: Ø Press [Power Meter] (1) To enter Relative mode (d. Br) Ø Press [Abs/Rel] (2) To return to Absolute mode (d. Bm): Ø Press [Abs/Rel] (3) 1 2 3 25
b. / Setting the Reference Must be in Relative mode d. Br not Absolute mode d. Bm Ø Press and hold soft button [Set Ref] for 3 seconds Ø Meter will beep 5 times Ø Autotest mode: meter will display ‘busy’ & zero at all incoming wavelengths Ø Manual mode: zero at indicated wavelength Ø Meter will not zero if display is ‘Lo’ Ø Referencing is retained at power off. [Set Ref] 26
c. / Log / Linear Mode To enable: 1 Ø Open hidden keypad Ø RS 232: press [d. Bm/w] (1) Ø USB: press [d. B/W] (2) 2 To return to d. B mode: Ø Repeat above key press • Must be in manual mode to switch between log & linear modes. • Linear display functions in both Manual & Autotest • Resets to log (d. Bm) mode at power off 27
8. / KI 7601 - Visible Laser To turn On: Ø Press [Source] (1) To toggle flashing: Ø Press [Mod] (2) 2 To turn Off: Ø Toggle [-/+] (3) 1 3 28
9. / Memory Operation a) Memory clear b) Memory store c) Memory store at a location d) Memory recall e) Memory extract to computer 29
a. / Memory clear Instrument must be in Manual mode, not Autotest Open hidden keypad: RS 232: Ø Press [CANCEL] and [RECALL MEMORY] simultaneously and hold for a few seconds. USB: Ø Press [MR] and [EXIT] simultaneously and hold for a few seconds. R S 2 3 2 U S B Meter will beep 5 times ‘clr’ will display when memory cleared. 30
a. / Memory - Store RS 232: Ø Open hidden keypad Ø Press [STORE] USB: Ø Press [M+] During Ø Ø Ø Store operation Display will blank or display BUSY Meter will beep once Memory location displayed top right side A full memory is indicated by a repeating buzzer In Autotest stores: All λs transmitted, Absolute Power, Reference value and S/N of the remote unit 31 M+
c. / Memory – RS 232 Store at a location 1 Exit Autotest Mode: Open hidden keypad Ø Ø 3 Press [RECALL MEMORY] (1) Toggle [-/+] to desired memory location (2) Press [SET] (3) When ready - Press [STORE] (4) 4 Note: memory writes continue from this location. Typical use: Ø Match memory location to fibre number 2 32
c. / Memory - USB Store at a location 3 1 Exit Autotest Mode: Open hidden keypad Ø Ø Press [MR] (1) Toggle [-/+] to desired memory location (2) Press [Set] (3) When ready - Press [M+] (4) 4 Note: memory writes continue from this location. Typical use: Ø Match memory location to fibre number 2 33
d. / Memory – Recall – RS 232 1 4 Exit Autotest Mode Open hidden keypad RS 232: Ø Push [RECALL MEMORY] (1) Ø Toggle [-/+] to desired memory location (2) Display alternates between λ and memory number Ø Push [ABS/REL] to alternate between d. Br & d. Bm (3) • d. Br: Reference shown on LHS of LCD. • Hint: Push & hold [ABS/REL] to display reference in display centre. Ø Press [-/+] to scroll λ and memory (2) Ø Exit memory display by pressing [CANCEL] (4) 2 34 3
d. / Memory – Recall – USB 4 Exit Autotest Mode Open hidden keypad 1 USB: Ø Push [MR] (1) Ø Toggle [-/+] to desired memory location (2) Display alternates between λ and memory number Ø Push [ABS/REL] to alternate between d. Br & d. Bm (3) • d. Br: Reference shown on LHS of LCD. • Hint: Push & hold [ABS/REL] to display reference in display centre. Ø Press [-/+] to scroll λ and memory (2) Ø Exit memory display by pressing [Exit] (4) 2 35 3
e. / Memory Extract to Computer Memory retrieve to computer is covered in the KITS™ training PPT Download to KITS software or to CSF file KITS: Familiar Excel™ user interface Lambd Readin Fiber a g Ref Remote S/N 1 1310 nm -7. 27 -8. 13 16549 1 1550 nm -7. 16 -7. 88 16549 2 1310 nm -7. 27 -8. 13 16549 2 1550 nm -7. 16 -7. 88 16549 3 1550 nm -60. 95 -7. 88 16549 4 1310 nm -39. 94 -8. 13 16549 4 1550 nm -30. 28 -7. 88 16549 5 1310 nm -39. 93 -8. 13 16549 5 1550 nm -30. 28 -7. 88 16549 6 1310 nm -43. 18 -8. 13 16549 6 1550 nm -41. 76 -7. 88 16549 36 CSV File: For those who do not use Microsoft Office