New N 1913 A and N 1914 A EPM Series Power Meters High speed New N 1913 A and N 1914 A EPM Series Power Meters High speed multi-channel measurement made possible! Agilent Restricted Page 1

Why should you consider migrating your existing E 4418/19 B to the new Agilent Why should you consider migrating your existing E 4418/19 B to the new Agilent N 1913/4 A? Existing platform: E 4418/9 B 1. Higher cost of ownership: • • • 2. Older platform with older user interface which make it harder to use. Lower number of reading/s Does not support new interfaces such as LAN and USB. Usage and application limitations: • • 5. Higher support cost System integration limitation • 4. Higher running cost: calibration on a yearly basis Longer operation time, higher operation cost: • • 3. More expensive at purchase. Only supports a limited set of power sensors. Only offer a maximum of 2 channels. Will no longer be orderable starting 1 Nov 2010. Agilent Restricted Page 2

What are the migration benefits? New platform: N 1913/4 A 1. Low cost of What are the migration benefits? New platform: N 1913/4 A 1. Low cost of ownership: • • • 2. 2 x Faster measurement speed Full support of GPIB, LAN/LXI, USB interfaces offering full integration flexibility. Support of a maximum number of power sensors. Up to 4 available channels to be used in various applications. Migration made easy • • 6. New intuitive easier to user interface. Versatile platform: • • 5. Lower support cost: up to 30% on up front support (3 year and 5 years) promotion 1. 056. System ready platform • 4. Lower running cost: calibration only once every 2 years. Shorter operation time, lower operation cost: • • 3. Cheaper at purchase. Backward code compatibility with the E 4418/9 B 20% trade discount when replacing the E 4418/9 B with the N 1913/4 A Leverage of all your previous power sensors investments: E 93 xx, 848 x, N 848 x… Financial solution to get easy and fast access to the new units: Easy Buy, Easy Rent, Easy Lease… www. microlease. com/agilent Already available for purchase Agilent Restricted Page 3

See the improvements for yourself! N 1913 A/14 A and E 4418 B/19 B See the improvements for yourself! N 1913 A/14 A and E 4418 B/19 B Outlook Comparison USB Host (U 2000 sensor channel) Color display VGA Out Trigger in/out Numerical keypad FRONT REAR USB, LAN and GPIB N 1913 A/14 A EPM series E 4418 B/19 B EPM series Agilent Restricted Page 4

New N 1913 A/14 A EPM Series Power Meters Replacing the Popular E 4418 New N 1913 A/14 A EPM Series Power Meters Replacing the Popular E 4418 B/19 B Power Meter High speed multi-channel measurement made possible! New N 1913 A/14 A provide: ü Faster measurement speed, up to 400 readings/sec 1 ü Up to 4 channels power measurements ü Support all Agilent average and CW sensors ü Code compatible with E 4418 B/19 B and legacy 436 A/7 B/8 A power meter 2 ü USB master, to work with U 2000 series USB sensor ü GPIB, LAN LXI & USB connectivity ü Smart Battery option – one push indicator function ü VGA output option ü Fast External Triggering option for power/frequency sweep function ü Software drivers for Labview (IVI-COM), Agilent VEE ü 2 years calibration interval 1. Comparing to E 4418 B/19 B using E-Series power sensors, 2. Need to purchase optional emulator software for 436 A/7 B/8 A Agilent Restricted Page 5

Product Briefing for N 1913 A/14 A and E 4418 B/19 B Specifications & Product Briefing for N 1913 A/14 A and E 4418 B/19 B Specifications & Features N 1913 A/14 A (1) E 4418 B/19 B Measurement Channel 1 or 2 -channel (additional 2 -channel via USB sensor ports, optional) 1 or 2 -channel Measurement speed 400 readings/s 200 readings/s Instrumentation Accuracy +/-0. 5% 1 m. W ref accuracy +/-0. 4% Calibration interval 2 years 1 year External Connectivity GPIB, LAN (LXI), USB GPIB, RS 232 Sensor support E-9300, E 4410, 8480, N 8480 series and U 2000 USB sensors E-9300, E 4410, 8480, N 8480 series Display Color Mono Data Input Device Numeric keypad Conventional 4 -key (up, down, left, right) Battery Life (optional) 7. 4 hrs (backlight off), 6 hrs (backlight off) Fully charged in 2. 2 hrs from empty stage 5. 5 hrs (backlight off), 3. 5 hrs (backlight on) Fully charged in 2 hrs from empty stage VGA, External Trigger In/Out (optional) Yes N/A Physical Dimension 212. 6 mm W x 88. 5 mm H x 348. 3 D (8. 5 in x 3. 5 in x 13. 7 in) Weight N 1913 A: 3. 6 kg (8. 2 lb) N 1914 A: 3. 7 kg (8. 4 lb) E 4418 B: 4. 0 kg (8. 8 lb) E 4419 B: 4. 1 kg (9. 0 lb) Agilent Restricted Page 6

N 1913 A/14 A Product Support Strategy • Support Warranty – Standard 1 year N 1913 A/14 A Product Support Strategy • Support Warranty – Standard 1 year warranty – Optional R-51 B-001 -3 C extended 3 years – Optional R-51 B-001 -3 C extended 5 years • Repair & Calibration – 2 years calibration interval (E 4418 B/19 B has 1 year calibration interval) – Repair & Calibration offered at WCSS centre, ready 8 weeks after SHIP – Upfront 1. 056 Full Service promotion • Support Parts at SPO – • Support Parts setup in SPO system for replacement and repair Migrate from E 4418 B/19 B Check out the migration guide : http: //cp. literature. agilent. com/litweb/pdf/5990 -4016 EN. pdf Agilent Restricted Page 7

Protect your investment with the widest range of power sensors support. Page 8 Protect your investment with the widest range of power sensors support. Page 8

Key benefits. Perform 3 -port or 4 -port measurement on a single power meter Key benefits. Perform 3 -port or 4 -port measurement on a single power meter Test case: Scalar Network Measurement on a power meter Improve test time up to 60% when stepping power, and 22% when stepping frequencies. Test case: Faster Power and Frequency Sweep Measurement Remotely control power measurement. Test case: Remote Power Monitoring Agilent Restricted Page 9

SCPI Commands Available in N 1913 A/14 A, Not in E 4418 B/19 B SCPI Commands Available in N 1913 A/14 A, Not in E 4418 B/19 B Migration guide : http: //cp. literature. agilent. com/litweb/pdf/5990 -4016 EN. pdf. N 1913/14 A CALCulate[1]|2|3|4: HOLD: STAT CALCulate[1]|2|3|4: FEED[1]|2 E 4418/19 B Not available Comments This command sets the hold value to be either minimum or maximum of the power measured. This is new feauture requirement to hold the CW/average power. This command sets the input measurement mode to be fed to the specified input on the CALC block. It is applied to the measurement after the CALC: MATH: EXPR command has been used to specify which channel the feed is taken from. Reserve for time gating function in future (USB sensor) CALibration 3|4: ZERO: TYPE EXTernal|INTernal OUTPut: RECorder[1]|2: FEED Not available This command is used to configure the U 2000 series power sensor either for external zeroing or internal zeroing. This command specifies which measurement is sent to the recorder output specified by the numeric value following RECorder. New feature for N 1913/14 A. Any channels can be assigned to become recorder A or B This command enables or disables the trigger output signal. OUTPut: TRIGger[: STATe] Not available E 4418/19 B does not have the Trigger output [SENSe[1]]|SENSe 2: BUFFer: COUNt [SENSe[1]]|2|3|4: CORRection: FDOFfset: UNIT Not available [SENSe[1]]|SENSe 2: FREQuency[: CW|: FIXed]: STARt Not available [SENSe[1]]|SENSe 2: FREQuency[: CW|: FIXEd]: STEP Not available [SENSe[1]]|SENSe 2: FREQuency[: CW|: FIXed]: STOP Not available This command sets the buffer size for average trigger measurement. It must be used in conjunction with external trigger. This command allows command query access to the Frequency Dependency Offset (FDO) unit New feature allow user to change the unit d. B or %. E 4418/19 B only allow in % This command sets the start frequency of frequency sweep. It must be used in conjunction with external trigger. Not available on E 4418/19 B This command sets the number of steps in frequency sweep. It must be used in conjunction with external trigger. Not available on E 4418/19 B This command sets the stop frequency of frequency sweep. It must be used in conjunction with external trigger. Not available on E 4418/19 B Agilent Restricted Page 10

SCPI Commands Available in N 1913 A/14 A, Not in E 4418 B/19 B SCPI Commands Available in N 1913 A/14 A, Not in E 4418 B/19 B N 1913/14 A E 4418/19 B Comments SYSTem: COMMunicate: LAN commands Not available SYSTem: DISPlay: BMP Not available SYSTem: HELP: HEADers? Not available SYSTem: LICense: INSTall Not available SYSTem: LICense: LIST? Not available SYSTem: LICense: REMove Not available This command will remove the valid installed option according to the license string. INITiate: CONTinuous: ALL Not available Sets all trigger sequences to be continuously initiated INITiate[: IMMediate]: ALL Not available TRIGger[: SEQuence[1]|2|3|4]: SLOPe Not available TRIGger[: SEQuence[1]|2|3|4]: COUNt Not available These control the settings for the LAN interface which is not available in E 4418/19 B This command returns the display image in bitmap format. Not available in E 4418/19 B. This command is limited to a maximum of five image returns per second. This query returns a list of all SCPI commands supported by the instrument. This command will install the valid licensed option according to the license string. Not available in E 4418/19 B. This query will return all the installed license string (for the options). Not available in E 4418/19 B. This command initiates all trigger sequences This new added feature in N 1913/14 A This command specifies whether a trigger event is recognized on the rising or falling edge of a signal (only on the etxernal trigger source). This command controls the path of the trigger subsystem in the upward traverse of the wait for trigger state. COUNt loops through the event detection/measurement cycle are performed. That is, COUNt measurements are performed in response to COUNt trigger events Applicable in FAST mode only. TRIGger[: SEQuence[1]|2|3|4]: DELay: AUTO TRIGger[: SEQuence[1]|2|3|4]: IMMedi ate Not available SERVice: BACKlight: BRIGhtness Not available SERVice: BACKlight: MODE Not available This command is used to determine whether or not there is a settling- time delay before a measurement is made. This command provides a one time over- ride of the normal process of the downward path through the wait for trigger state. Trigger source on bus or hold only This command sets the intensity of front panel’s backlight. E 4418/19 B does not have this feature on the LCD control. This command sets the mode of front panel’s backlight to ON, OFF or TIMED. When TIMED mode is selected, the backlight will turn off within 10 minutes after the last keypress. Press any keys on the front panel to turn the backlight on again. Agilent Restricted Page 11

SCPI Commands Available in N 1913 A/14 A, Not in E 4418 B/19 B SCPI Commands Available in N 1913 A/14 A, Not in E 4418 B/19 B N 1913/14 A E 4418/19 B SERVice: BIST: commands Not available SERVice: CALibrator: ADJ: commands Not available SERVice: DISPlay: VGA Comments These command enables/disables the Built In Self Test (BIST) for calibrator, linearity and zero set tests. This is new added feature in N 1913/14 A. E 4418/19 runs the builtin test during power up These command adjust the 1 m. W calibrator output in coarse scale or fine scale This command enables or disables the state of VGA output Not available SERVice: DISPlay: BSCReen commands Not available SERVice: FAN: FULL Not available SERVice: LAN: PHOStname Not available SERVice: SECure: ERASe Not available SERVice: SENSor[1]|2: CALFactor Not available SERVice: SENSor[1]|2|3|4: FREQuency commands Not available SERVice: SENSor[1]|2|3|4: POWer: AV ERage: MAXimum? SERVice: SENSor[1]|2|3|4: POWer: US ABle commands SERVice: SENSor[1]|2|3|4: RADC? Not available This is new added feature in N 1913/14 A for the VGA option. This command enables or disables the Blank Screen and Secure Blank Screen features of the front panel display. This command set the fan controller to run at full speed. At the first power- up, the default state is ON. If the user sets the state to OFF and power cycle, the state will be saved. On the next power cycle, the state will be set to OFF. This command preset the LAN hostname to its default value. It requires the serial number to be set- up. This command erases the EPM Series power meter’s memory, for example, before you return it to Agilent Technologies for repair or calibration, of all data stored in it. This command writes calibration factor data to, or reads calibration factor data from, the currently connected sensor. The whole calibration factor block must be written at once as a checksum is generated. The new block must not be larger than the existing block. This command applies to the following sensors: • E 4410 Series • N 8480 Series (excluding Option CFT) • E 9300 Series These query returns the maximum/minimum frequency that can be measured by the currently connected sensor. It is applicable to E- Series sensors only. Maximum frequency information is stored in the sensor’s EEPROM This query returns the maximum average power that can be measured by the currently connected sensor. It is applicable to E- Series sensors only Not available returns the maximum/minimum power that can be accurately measured by the currently connected sensor. Not available This query returns a new raw uncorrected measurement in volts, as a 32 bit signed integer This query returns the tracking number for U 2000 Series and E- Series sensors. Tracking number information is stored in the sensor’s EEPROM This command enables or disables the warm start feature SERVice: SENSor[1]|2|3|4: TNUMber? Not available SERVice: STATe Not available Agilent Restricted Page 12

SCPI Commands Available in E 4418 B/19 B, Not in N 1913 A/14 A SCPI Commands Available in E 4418 B/19 B, Not in N 1913 A/14 A N 1913/14 A E 4418/19 B Not available CALibration[1|2]: ECONtrol: STATe Comments This command is used to enable and disable the rear panel TTL zero/cal inputs. The TTL inputs provide an external means to initiate ZERO and CAL cycles. N 1913/14 A does not have the remote TTL I/O This command is used to control the display contrast. Not available DISPlay: CONTrast OUTPut: TTL[1|2]: FEED N 1913/14 A Color LCD does not contrast control, only backlight control This command controls whether a window limits fail drives the rear panel TTL Output HIGH or LOW. N 1913/14 A does not have the TTL out limit function Not available OUTPut: TTL[1|2]: FEED This command controls whether an asserted rear panel TTL Output indicates a failure in the upper, lower or either limit for the chosen window. Not available OUTPut: TTL[1|2]: STATe This command is used to enable and disable the rear panel Upper and Lower window TTL Outputs. Not available [SENSe[1]]|SENSe 2: LIMit: CLEar commands Other [SENSe[1]]|SENSe 2: LIMit commands Not available SYStem: COMMunicate: Serial commands Not available SYSTem: RINTerface GPIB|RS 232|RS 422 Not available SERVice: OPTion used to clear any limit failures in E 4418/19 B. Not supported in N 1913/14 A This feature is not supported in N 1913/14 A These control the settings for the RS 232/RS 422 serial interface which is not available in N 1913/14 A This command allows the remote control interface to be selected from GPIB, RS 232, and RS 422. N 1913/14 A does not need to select the interfaces , will auto detect & configure which interface to use This command loads the power meter memory with the options fitted. The query form of the command can be used to determine which options are fitted to the unit. For N 1913/14 A, user can use the *opt? query This command is used to load the power meter with a serial number in the form GB 12345678 or US 12345678. Not available SERVice: SNUMber For N 1913/14 A, user can only query the serial number: SERVice: SENSor[1]|2|3|4: SNUMber? Agilent Restricted Page 13

Resources for N 1913 A/14 A QFS: http: //field-bid. marketing. agilent. com/file_library/N 1913_4 A%20 Resources for N 1913 A/14 A QFS: http: //field-bid. marketing. agilent. com/file_library/N 1913_4 A%20 QFS_direct. pdf PMPS Brochure: http: //cp. literature. agilent. com/litweb/pdf/5989 -6240 EN. pdf PMPS DM: http: //cp. literature. agilent. com/litweb/pdf/5989 -7837 EN. pdf Data sheet: http: //cp. literature. agilent. com/litweb/pdf/5990 -4019 EN. pdf User’s guide: http: //cp. literature. agilent. com/litweb/pdf/N 1913 -90001. pdf Programmer’s guide: http: //cp. literature. agilent. com/litweb/pdf/N 1913 -90008. pdf Installation guide: http: //cp. literature. agilent. com/litweb/pdf/N 1913 -90010. pdf Service guide: http: //cp. literature. agilent. com/litweb/pdf/N 1913 -90009. pdf Migration guide : http: //cp. literature. agilent. com/litweb/pdf/5990 -4016 EN. pdf. Where to Find the Details? • Website: www. agilent. com/find/epm For support: Contact your local Agilent office http: //www. home. agilent. com/agilent/contact. Information. jspx? cc=US&lc=eng&nid=11143. 0&page. Mode=HB&tmprop=TM Agilent Restricted Page 14

Product Options Model / Option Description N 1913 A/N 1914 A Single/dual channel average Product Options Model / Option Description N 1913 A/N 1914 A Single/dual channel average power meter N 1913 A/14 A-102 Single/dual channel average power meter, battery N 1913 A/14 A-103 Single/dual channel average power meter, battery, 2 rears USB ports N 1913 A/14 A-104 Single/dual channel average power meter, VGA N 1913 A/14 A-105 Single/dual channel average power meter, battery, 1 front 1 rear USB ports, VGA N 1913 A/14 A-106 Single/dual channel average power meter, 1 front 1 rear USB ports N 1913 A/14 A-107 Single/dual channel average power meter, external trigger in/out N 1913 A/14 A-108 Parallel rear panel sensor input connectors and front panel reference calibrator connector, 2 rear USB ports N 1913 A/14 A-109 Parallel rear panel sensor input connectors and rear panel reference calibrator connector, 2 rear USB ports N 1913 A/14 A-200 436 A & 437 B code compatibility Agilent Restricted Page 15

E 4418 B and E 4419 B Discontinue Plan E 4418 B/19 B Disco E 4418 B and E 4419 B Discontinue Plan E 4418 B/19 B Disco Start 2 Aug 2010 MAY 2009 JUN 2009 JUL Launch N 1913 A/14 A (9 Jun 09) … … AUG 2010 E 4418 B/19 B Last Order Date 1 Nov 2010 SEP OCT NOV DEC E 4418 B/19 B Last Ship Date 1 Feb 2011 JAN 2011 FEB E 4418 B/19 B Obsolete 2 Nov 2015 … DEC 2011 2012 2013 2014 2015 2016 2017 5 Years Support Life Agilent Restricted Page 16

Why Choose the New N 1913 A/14 A? High speed multi-channel measurement made possible! Why Choose the New N 1913 A/14 A? High speed multi-channel measurement made possible! 1. Protect your long term investment: • • Backward compatible with E 4418 B/19 B, 436 A, 437 B and 438 A power meters Widest power sensors support (compatible with all Agilent CW and E-Series power sensors) 2. Reduce time to market: • • 2 times faster than E 4418 B/19 B. Calibrate only once every 2 years. • • Versatile platform. Can be used for many different applications. Trade in your old E 4418 B, E 4419 B, 435 B, 436 A, 437 B, 438 A, 8990 A and 8991 A power meters and receive 20% discount on your N 1913 A and N 1914 A purchase. Find out more: www. agilent. com/find/pmpromo. Upfront 1. 056 Full Service promotion Financial solutions: Easy Buy, Easy Rent, Easy Lease… www. microlease. com/agilent 3. Reduce cost of ownership • • Agilent Restricted Page 17

Q and A Session THANK YOU Agilent Restricted Page 18 Q and A Session THANK YOU Agilent Restricted Page 18