Low-Temperature Sterilization: Characteristics, Benefits, Monitoring William A. Rutala, Ph. D, MPH Director, Hospital Epidemiology, Low-Temperature Sterilization: Characteristics, Benefits, Monitoring William A. Rutala, Ph. D, MPH Director, Hospital Epidemiology, Occupational Health and Safety at UNC Health Care System; Professor of Medicine and Director, Statewide Program for Infection Control and Epidemiology at University of North Carolina at Chapel Hill, USA Disclosure-3 M

Low-Temperature Sterilization Ä Provide overview of low-temperature sterilization n Characteristics n Benefits n Monitoring Low-Temperature Sterilization Ä Provide overview of low-temperature sterilization n Characteristics n Benefits n Monitoring

www. disinfectionandsterilizatio n. org www. disinfectionandsterilizatio n. org

Disinfection and Sterilization in Healthcare Facilities WA Rutala, DJ Weber, and HICPAC, www. cdc. Disinfection and Sterilization in Healthcare Facilities WA Rutala, DJ Weber, and HICPAC, www. cdc. gov l Overview Last Centers for Disease Control and Prevention guideline in 1985 n 158 pages (>82 pages preamble, 34 pages recommendations, glossary of terms, tables/figures, >1000 references) n Evidence-based guideline n Cleared by HICPAC February 2003; delayed by FDA n Published in November 2008 n

Disinfection and Sterilization EH Spaulding believed that how an object will be disinfected depended Disinfection and Sterilization EH Spaulding believed that how an object will be disinfected depended on the object’s intended use. CRITICAL - objects which enter normally sterile tissue or the vascular system or through which blood flows should be sterile. SEMICRITICAL - objects that touch mucous membranes or skin that is not intact require a disinfection process (high-level disinfection [HLD]) that kills all microorganisms but high numbers of bacterial spores. NONCRITICAL -objects that touch only intact skin

Efficacy of Sterilization Influencing Factors Cleaning of the object Organic and inorganic load present Efficacy of Sterilization Influencing Factors Cleaning of the object Organic and inorganic load present Type and level of microbial contamination Concentration of and exposure time to disinfectant/sterilant Nature of the object Temperature and relative humidity

Processing “Critical” Patient Care Objects Classification: Critical objects enter normally sterile tissue or vascular Processing “Critical” Patient Care Objects Classification: Critical objects enter normally sterile tissue or vascular system, or through which blood flows. Object: Sterility. Level germicidal action: Kill all microorganisms, including bacterial spores. Examples: Surgical instruments and devices; cardiac catheters; implants; etc. Method: Steam, gas, hydrogen peroxide gas plasma, vaporized hydrogen peroxide, ozone or chemical sterilization.

Critical Objects l Surgical instruments l Cardiac catheters l Implants Critical Objects l Surgical instruments l Cardiac catheters l Implants

Sterilization The complete elimination or destruction of all forms of microbial life and is Sterilization The complete elimination or destruction of all forms of microbial life and is accomplished in healthcare facilities by either physical or chemical processes

Sterilization of Patient Equipment Alternatives to ETO-CFC ETO-CO 2, ETO-HCFC, 100% ETO l Newer Sterilization of Patient Equipment Alternatives to ETO-CFC ETO-CO 2, ETO-HCFC, 100% ETO l Newer Low Temperature Sterilization Technology Hydrogen Peroxide Gas Plasma Vaporized hydrogen peroxide Ozone-not commonly used l

Cleaning l Mechanical cleaning machines-automated equipment may increase productivity, improve cleaning effectiveness, and decrease Cleaning l Mechanical cleaning machines-automated equipment may increase productivity, improve cleaning effectiveness, and decrease worker exposure n n n l Utensil washer-sanitizer Ultrasonic cleaner Washer sterilizer Dishwasher Washer disinfector Manual

STERILIZATION Factors affecting the efficacy of sterilization l l l l Bioburden Cleaning Pathogen STERILIZATION Factors affecting the efficacy of sterilization l l l l Bioburden Cleaning Pathogen type Protein and salt Biofilm accumulation Lumen length and diameter Restricted flow

Lumens Sterilized by Various Low-Temperature Sterilization Methods Challenge: 12/88 100%ETO HCFC-ETO Sterrad 10% Serum, Lumens Sterilized by Various Low-Temperature Sterilization Methods Challenge: 12/88 100%ETO HCFC-ETO Sterrad 10% Serum, 0. 65% Salt (7 organisms, N=63) 44% 39. 7% 49. 2% 35% No Serum or Salt, (3 organisms, N=27) ND 96. 3% ND Alfa et al. Infect Cont Hosp Epidemiol 1996; 17: 92 -100. The three organisms included: E. faecalis, M. chelonei, B. subtilis spores. The seven organisms included: E. faecalis, P. aeruginosa, E. coli, M. chelonei, B. subtilis spores, B. stearothermophilus

Low-Temperature Sterilization Technologies (LTST) Conclusions l All LTST technologies have limitations l LTST (ETO, Low-Temperature Sterilization Technologies (LTST) Conclusions l All LTST technologies have limitations l LTST (ETO, HP gas plasma) demonstrate a significant number of failures in presence of serum or salt l Salt and serum provide protection for spores and bacteria l Salt and serum combined with a narrow lumen provide extraordinary protection with LTST

IS THERE A STANDARD TO DEFINE WHEN A DEVICE IS CLEAN? l l There IS THERE A STANDARD TO DEFINE WHEN A DEVICE IS CLEAN? l l There is currently no standard to define when a device is “clean”, cleanliness controlled by visual Potential methods: level of detectable bacteria; protein (6µg/cm 2); endotoxin; ATP; lipid This is due in part to the fact that no universally accepted test soils to evaluate cleaning efficiency and no standard procedure for measuring cleaning efficiency At a minimum, a cleaning process should: reduce the natural bioburden; remove organic/inorganic

Verifying Cleaning Processes AAMI recommends incorporating test methods that verify the functionality of the Verifying Cleaning Processes AAMI recommends incorporating test methods that verify the functionality of the automated washer l Washer indicators have been in use in Europe and Canada and some US hospitals l Washer indicators are chemical indicators imprinted with a dried test soil formula and a dye. l

Cleaning Indicators for Washers l l Monitor the automated washer and instrument cleaning chemistry Cleaning Indicators for Washers l l Monitor the automated washer and instrument cleaning chemistry functionaiity Indicator includes proteins, lipids, and polysaccharides to mimic common challenging test soils

Low-Temperature Sterilization l Provide overview of low-temperature sterilization n Characteristics n Benefits-sterilize temperature and Low-Temperature Sterilization l Provide overview of low-temperature sterilization n Characteristics n Benefits-sterilize temperature and moisturesensitive medical and surgical devices in health care u. Review advantages/disadvantages of LTS technology n Monitoring

100% Ethylene Oxide 4. 8 ft 3 chamber 100% Ethylene Oxide 4. 8 ft 3 chamber

Ethylene Oxide (ETO) l l Advantages n Very effective at killing microorganisms n Penetrates Ethylene Oxide (ETO) l l Advantages n Very effective at killing microorganisms n Penetrates medical packaging and many plastics n Compatible with most medical materials n Cycle easy to control and monitor Disadvantages n Some states (CA, NY, TX) require ETO emission reduction of 90 -99. 9% (air pollution control device) n CFC (inert gas that eliminates explosion hazard) banned after 1995 n Potential hazard to patients and staff n Lengthy cycle/aeration time

Ethylene Oxide Aeration • The ability to penetrate is an important characteristic of ETO Ethylene Oxide Aeration • The ability to penetrate is an important characteristic of ETO • It is this property that necessitates extended periods of aeration • During aeration, heated air is flushed through chamber to remove ETO residuals from sterilized items • Consult the medical device manufacturer for recommended aeration times • Composition of device

Ethylene Oxide Aeration Time Guidelines* Temperature Aeration Time 50°C 12 hours 60°C 8 hours Ethylene Oxide Aeration Time Guidelines* Temperature Aeration Time 50°C 12 hours 60°C 8 hours * Based on polyvinylchloride (PVC) tubing Ref: ANSI/AAMI ST 41: 2008 Section 8. 8. 4

OSHA Occupational Exposure Standard for ETO Exposure Limits Employer required to monitor employee exposure OSHA Occupational Exposure Standard for ETO Exposure Limits Employer required to monitor employee exposure WHAT? • Monitor for 8 -hour TWA • 15 minute sampling period HOW? 3 M™ Ethylene Oxide Monitor 3550 • Breathing zone air samples WHO? • Representative employee on each shift for each job classification If ETO exposure 0. 5 ppm Action Level, 8 -hr TWA-monitoring of those employees can be discontinued

Hydrogen Peroxide Gas Plasma Hydrogen Peroxide Gas Plasma

Hydrogen Peroxide Gas Plasma Sterilization Advantages l Safe for the environment and health care Hydrogen Peroxide Gas Plasma Sterilization Advantages l Safe for the environment and health care worker; it leaves no toxic residuals l Fast - cycle time is 28 -52 min and no aeration necessary l Used for heat and moisture sensitive items since process temperature 50 o. C l Simple to operate, install, and monitor l Compatible with most medical devices

Hydrogen Peroxide Gas Plasma Sterilization Disadvantages l Cellulose (paper), linens and liquids cannot be Hydrogen Peroxide Gas Plasma Sterilization Disadvantages l Cellulose (paper), linens and liquids cannot be processed l Sterilization chamber , about 3. 5 ft 3 to 7. 3 ft 3 l Endoscopes or medical devices restrictions based on lumen internal diameter and length (see manufacturer’s recommendations); expanded claims with NX l Requires synthetic packaging (polypropylene) and special container tray

V-PRO™ 1, Vaporized Hydrogen Peroxide l Advantages n n n l Safe for the V-PRO™ 1, Vaporized Hydrogen Peroxide l Advantages n n n l Safe for the environment and health care worker; it leaves no toxic residuals Fast - cycle time is 55 min and no aeration necessary Used for heat and moisture sensitive items (metal and nonmetal devices) Disadvantages n n n Sterilization chamber is small, about 4. 8 ft 3 Medical devices restrictions based on lumen internal diameter and length-see manufacturer’s recommendations, e. g. , SS lumen 1 mm diameter, 125 mm length Not used for liquid, linens, powders, or any cellulose materials Requires synthetic packaging (polypropylene) Limited use and limited comparative microbicidal efficacy

Ozone l Advantages n n l Used for moisture and heat-sensitive items Ozone generated Ozone l Advantages n n l Used for moisture and heat-sensitive items Ozone generated from oxygen and water (oxidizing) No aeration because no toxic by-products FDA cleared for metal and plastic surgical instruments, including some instruments with lumens Disadvantages n n Sterilization chamber small, 4 ft 3 Limited use (material compatibility/penetrability/organic material resistance? ) and limited microbicidal efficacy data

Conclusions Sterilization All sterilization processes effective in killing spores l Cleaning removes salts and Conclusions Sterilization All sterilization processes effective in killing spores l Cleaning removes salts and proteins and must precede sterilization l Failure to clean or ensure exposure of microorganisms to sterilant (e. g. connectors) could affect effectiveness of sterilization process l

Recommendations Methods of Sterilization Steam is preferred for critical items not damaged by heat Recommendations Methods of Sterilization Steam is preferred for critical items not damaged by heat l Follow the operating parameters recommended by the manufacturer (times, temperatures, gas conc) l Use low temperature sterilization technologies for reprocessing critical items damaged by heat l Aerate surgical and medical items that have been sterilized in the ETO sterilizer l

Newer Trends in Sterilization of Patient Equipment l Alternatives to ETO-CFC ETO-CO 2, ETO-HCFC, Newer Trends in Sterilization of Patient Equipment l Alternatives to ETO-CFC ETO-CO 2, ETO-HCFC, 100% ETO l New Low Temperature Sterilization Technology Hydrogen peroxide gas plasma-most common Vaporized hydrogen peroxide-limited clinical use Ozone and hydrogen peroxide-not FDA cleared Nitrogen dioxide-not FDA cleared

Low-Temperature Sterilization l Provide overview of low-temperature sterilization n Characteristics n Benefits-sterilize temperature and Low-Temperature Sterilization l Provide overview of low-temperature sterilization n Characteristics n Benefits-sterilize temperature and moisturesensitive medical and surgical devices in health care n Monitoring

Monitoring of Sterilizers Use physical, chemical and biological monitors to ensure the effectiveness of Monitoring of Sterilizers Use physical, chemical and biological monitors to ensure the effectiveness of the sterilization process l Each load should be monitored with physical and chemical indicators l If the physical, chemical or biological indicators suggest inadequate processing, the items should not be used l Biological indicators should be used at least l

Objectives of Monitoring the Sterilization Process l Assures probability of absence of all living Objectives of Monitoring the Sterilization Process l Assures probability of absence of all living organisms on medical devices being processed l Detect failures as soon as possible l Removes medical device involved in failures before patient use

Sterilization Monitoring Sterilization monitored routinely by combination of physical, chemical, and biological parameters l Sterilization Monitoring Sterilization monitored routinely by combination of physical, chemical, and biological parameters l Physical - cycle time, temperature, pressure l Chemical - heat or chemical sensitive inks that change color when germicidal-related parameters present (Class 1 -6) l Biological - Bacillus spores that directly measure sterilization

Biological Indicators contain spores • Select BIs that of Bacillus atrophaeus • Rationale: BIs Biological Indicators contain spores • Select BIs that of Bacillus atrophaeus • Rationale: BIs are the only sterilization process monitoring device that provide a direct measure of the lethality of the process

Biological Monitors ETO - B. atrophaeus l New low temperature sterilization technologies HP gas Biological Monitors ETO - B. atrophaeus l New low temperature sterilization technologies HP gas plasma (Sterrad) and HP vapor (VPro)- G. stearothermophilus Ozone-G. stearothermophilus l

Process Challenge Device (PCD) “A PCD is a device used to assess the effective Process Challenge Device (PCD) “A PCD is a device used to assess the effective performance of a sterilization process by providing a challenge to the process that is equal or greater than the challenge posed by the most difficult item routinely processed. ” • PCD may be: • User-assembled test pack • Commercially available, FDA cleared preassembled test pack • ST 41 describes two types of PCDs • Routine BI test pack

Ethylene Oxide Sterilization Recommended Monitoring Practices Routine Load Release/Sterilizer Efficacy Monitoring • Physical monitoring Ethylene Oxide Sterilization Recommended Monitoring Practices Routine Load Release/Sterilizer Efficacy Monitoring • Physical monitoring of cycle • External and internal monitoring of packages • Monitoring of every load with a PCD (routine test pack) containing a BI and a CI or equivalent, commercially available BI PCD ANSI/AAMI ST 41: 2008, Section 10 • Ideally, quarantine implants until BI results

Ethylene Oxide BI placed in full load in center l Frequency, in each load Ethylene Oxide BI placed in full load in center l Frequency, in each load l Visual, color 48 hour readout l Rapid readout BI for ETO provides 4 hr l

Hydrogen Peroxide Gas Plasma l l Provides evidence that proper sterilization conditions were achieved Hydrogen Peroxide Gas Plasma l l Provides evidence that proper sterilization conditions were achieved in 24 hours Built-in chemical indicator confirms it has been exposed to HP if color changed from red to yellow Placed in peel pack on bottom rack with tyvek side up 1 x per day CI placed in every load

Monitoring of Sterilizers l Internal Chemical Indicator n Validates the sterilant penetrated the pack Monitoring of Sterilizers l Internal Chemical Indicator n Validates the sterilant penetrated the pack or tray n Advantage of the pack control monitor is that it is inside each pack in multiple locations n Detect local problem

Recommendations Monitoring of Sterilizers l Following a single positive biological indicator used with a Recommendations Monitoring of Sterilizers l Following a single positive biological indicator used with a method other than steam, treat as non-sterile all items that have been processed in that sterilizer, dating back to last negative biological indicator. These non-sterile items should be retrieved, if possible, and reprocessed.

Routine Load Release/Routine Sterilizer Efficacy Monitoring Acceptance criteria • Appropriate readings from physical monitors Routine Load Release/Routine Sterilizer Efficacy Monitoring Acceptance criteria • Appropriate readings from physical monitors • Appropriate readings from CIs • Negative result from BI in PCD • Positive BI control result • Run a control BI, with matching lot #, each day • Document all results in sterilization cycle record ANSI/AAMI ST 41: 2008, Section 10

Recommendations Storage of Sterile Items Sterile storage area should be well-ventilated area that provides Recommendations Storage of Sterile Items Sterile storage area should be well-ventilated area that provides protection against dust, moisture, and temperature and humidity extremes. l Sterile items should be stored so that packaging is not compromised l Sterilized items should be labeled with a load number that indicates the sterilizer used, the cycle or load number, the date of sterilization, l

Conclusions All sterilization processes effective in killing spores l Cleaning removes salts and proteins Conclusions All sterilization processes effective in killing spores l Cleaning removes salts and proteins and must precede sterilization l Delivery of sterile products for use in patient care depends not only on the effectiveness of the sterilization process but also on cleaning, disassembling and packaging of the device, loading the sterilizer, and monitoring l

Low-Temperature Sterilization l Provide overview of low-temperature sterilization n Characteristics n Benefits n Monitoring Low-Temperature Sterilization l Provide overview of low-temperature sterilization n Characteristics n Benefits n Monitoring

References l l Rutala WA, Weber DJ. Disinfection and sterilization: What clinicians need to References l l Rutala WA, Weber DJ. Disinfection and sterilization: What clinicians need to know. Clin Infect Dis 2004; 39: 702 Rutala WA, Weber DJ, HICPAC. CDC guideline for disinfection and sterilization in healthcare facilities. November 2008. www. cdc. gov Alfa MJ, Olson N, De. Gagne P, Hizon R. New low temperature sterilization technologies: Microbicidal activity and clinical efficiency. In Rutala WA, ed. Disinfection, Sterilization, and Antisepsis in Healthcare. Champlain, NY: Polyscience Publications. 1998: 67 -78. Rutala WA, Weber DJ. Clinical effectiveness of lowtemperature sterilization technologies. Infect Control Hosp