Скачать презентацию FLUROSCOPY QUALITY CONTROL A fluoroscopy system monitored Скачать презентацию FLUROSCOPY QUALITY CONTROL A fluoroscopy system monitored

89ece928c0df0d7ca342997969fe3755.ppt

  • Количество слайдов: 33

FLUROSCOPY QUALITY CONTROL FLUROSCOPY QUALITY CONTROL

A fluoroscopy system monitored to achieve good image quality l l l The automatic A fluoroscopy system monitored to achieve good image quality l l l The automatic exposure rate control (AEC) and image quality are important aspects of a fluoroscopy system. Spatial resolution is measured with a contrast-detail test object is used to measure threshold contrast-detail performance. Median limiting spatial resolution was 1. 41 lp. mm-1 for image intensifier fields lying between 18 and 24 cm in diameter.

l l FLUORO OUTPUT MEASURED WITH TYPE OF PHANTOM MONITORED? RECORDED ? l l FLUORO OUTPUT MEASURED WITH TYPE OF PHANTOM MONITORED? RECORDED ?

FLUOROSCOPY QUALITY CONTROL: l l MONIOTRED – WEEKLY WITH A PHANTOM 7 IN LUCITE FLUOROSCOPY QUALITY CONTROL: l l MONIOTRED – WEEKLY WITH A PHANTOM 7 IN LUCITE OF 9” WATER PLACED IN PATH OF BEAM – OUTPUT RECORDED OUTPUT (WITH PATIENTS ) DONE DAILY

l l WEAR PROTECTIVE GARMENTS while performing this procedure! Activate the fluoro beam and l l WEAR PROTECTIVE GARMENTS while performing this procedure! Activate the fluoro beam and move the fluoro tower as required in order to center the phantom on Cone down (close) the collimators until only edges of the phantom are visible.

Phantom Images l l l To ensure that film density, contrast, uniformity, and image Phantom Images l l l To ensure that film density, contrast, uniformity, and image quality due to the x-ray imaging system and film processor are maintained at optimum levels. When imaging the phantom for fluoroscopic tests, record the k. Vp and m. A from the meters or displays provided. Any changes in the k. Vp or m. A values for a constant phantom thickness can indicate problems with the fluoroscopic chain, for example, a loss of image intensifier/TV gain.

Digital fluoro resolution test Digital fluoro resolution test

CONSISTENCY IS THE KEY l l After each link (x-ray unit, video monitor, laser CONSISTENCY IS THE KEY l l After each link (x-ray unit, video monitor, laser film printer, processor, darkroom, etc. ) in the imaging chain is optimized, a working QA program will provide warning flags to the QA program coordinator when something goes awry. the test or tests must be repeated to verify the results, then corrective action must be taken. The coordinator must be capable of identifying problems and willing to resolve them as they occur, or the QA program will not provide the intended benefits.

Equipment Warm-up DAILY l l l Each day during the x-ray generator warmup, and Equipment Warm-up DAILY l l l Each day during the x-ray generator warmup, and before exposing the first patient to xrays, check for indicator malfunction (m. A, fiveminute timer, etc. ) and the mechanical and electrical safety of the x-ray system. Malfunctions and unsafe conditions (such as frayed wires) must be corrected promptly.

Equipment Warm-up DAILY l l l NOTE: because this procedure necessitates activation of the Equipment Warm-up DAILY l l l NOTE: because this procedure necessitates activation of the fluoroscopic tube, the surveyor should make certain there is something (phantom, cubic water container, etc. ) attenuating the useful x-ray beam during operation and wear appropriate protective equipment (lead apron, gloves, thyroid shield, etc. )

FOR DIGITAL SYSTEMS: l l l . Verify that the window and level settings FOR DIGITAL SYSTEMS: l l l . Verify that the window and level settings are set to the pre- established levels. Adjust room lighting conditions to those used clinically. Observe the mesh patterns, line pairs or other test objects under fluoroscopy

Test Frequency – Weekly l l l Laser Film Printer Quality Control - SMPTE Test Frequency – Weekly l l l Laser Film Printer Quality Control - SMPTE Test Pattern Video Monitor Setup =Must be performed on EACH video monitor (operator’s console, physician’s console, etc. ) so that the grays on the monitors all appear similar. Laser Printer Setup 1. Send the SMPTE test pattern to the laser film printer using the largest most commonly used image format (1 on 1 or 4 on 1). 2. Process film if applicable. 3. With the film on a view box and the same image on the monitor, visually compare the film gray scale densities on the film to those on the monitor.

SMPTE Test Pattern ? l l l Society of Motion Picture and Television Engineers SMPTE Test Pattern ? l l l Society of Motion Picture and Television Engineers

Monitor Quality l l l Often the monitor is the weakest link in the Monitor Quality l l l Often the monitor is the weakest link in the digital imaging chain. The monitor has a direct effect on the quality of the image that is presented to the radiologist for reading or to the referring physician for review. Not every computer that is used to view images can have the highest-quality monitor available because of the expense.

Monitor Quality l l l Radiologist’s workstation has highest-quality monitor, usually 2 K or Monitor Quality l l l Radiologist’s workstation has highest-quality monitor, usually 2 K or 3 K for computed radiography or digital radiography, 1 K or 2 K for cross-sectional images, and up to 4 K for mammography. Megapixel measurement may be used to determine the appropriate monitor. Generally, physician’s review workstations and technologist’s QC workstations have high-quality commercial monitors. • Resolution is usually 1 K.

Wet Laser Imager l Daily/weekly QC: • • Monitor each film printed to ensure Wet Laser Imager l Daily/weekly QC: • • Monitor each film printed to ensure that it is free from artifacts and matches monitor or desired quality. Print a test pattern from the PACS, or some printers have a built-in test pattern that can be printed by depressing a button on the printer.

VIGNETTING PINCUSION EFFECT The center has the best resolution/detail VIGNETTING PINCUSION EFFECT The center has the best resolution/detail

DICOM gray scale function window level and width function l A photometer to a DICOM gray scale function window level and width function l A photometer to a monitor screen in a check of the monitor's conformance with the DICOM Grayscale Standard Display Function.

PACS QC Activities l ACR also suggests that all of the following tests be PACS QC Activities l ACR also suggests that all of the following tests be carried out with a SMPTE (Society of Motion Pictures and Television Engineers) test pattern to ensure continuity of measurements.

Equipment Visual Checklist Form l l l All foot and hand switches designed to Equipment Visual Checklist Form l l l All foot and hand switches designed to energize the fluoroscopic tube should be tested to ensure that x-ray production is terminated as soon as the switch is released. If switches have multiple positions (example: high level control) each position should be tested. The table, image intensifier, and tube, as applicable, should move smoothly and freely without requiring excessive force. Check all of the locks and centering detents on the tower and table for adequate function. Check all switches, lights and meters on the table, control panel and image intensifier for appropriate function. (i. e. lights should light, meters should move).

Equipment Visual Checklist Form l l l Lead drapes should be affixed to the Equipment Visual Checklist Form l l l Lead drapes should be affixed to the image intensifier (under table systems) and have no cracks Bucky moves smoothly along the track and its locking mechanism is functioning. Bucky slot during fluoroscopy is working as designed.

Equipment Visual Checklist Form l l l If the unit is a portable c-arm, Equipment Visual Checklist Form l l l If the unit is a portable c-arm, it must be equipped with a spacer on the tube to prevent the patient from being closer than 30 cm to the tube’s target. This spacer must always be used unless it interferes with a sterile field as during surgery. It must not be possible to activate the x-ray tube unless the entire fluoroscopic beam is intercepted by the image receptor. On systems where the image intensifier can be placed in a park position, do so and step on the pedal to assure that this interlock is functioning. Verify that all of the indicators on the control panel are functioning. Check that the fluoro timer emits an audible signal or terminates the exposure after 5 minutes expires.

Lead Aprons - annually l l Examine the integrity of the personnel shielding devices Lead Aprons - annually l l Examine the integrity of the personnel shielding devices to ensure optimal protection to the patient when positioned properly. NOTE: Lead aprons should never be folded. Cracks in the lead lining can develop at the fold, reducing the useful life of the apron. Do not assume that brand new aprons, gloves, etc. contain no defects. New aprons, gloves, etc. should be examined under x- ray immediately upon arrival and returned to supplier if defects are found.

Proper storage for apron l l Hang or lay flat Clean with a damp Proper storage for apron l l Hang or lay flat Clean with a damp cloth Cannot be laundered DO NOT FOLD

Fluoroscopic Systems l Intensifier viewing system resolution Intensifier viewing system contrast TV monitors and Fluoroscopic Systems l Intensifier viewing system resolution Intensifier viewing system contrast TV monitors and recorders Automatic brightness control l LEAD APRONS - CRACKS l l l

Output, Scatter, Leakage Check shielding Output, Scatter, Leakage Check shielding

OVERFRAMING Collimation check OVERFRAMING Collimation check

System Training l l l Each department also has a list of skills that System Training l l l Each department also has a list of skills that are tested and retrained each year. PACS skills should also be included in this annual training. A training record should be kept for each employee to show proof of the skills in which the employee has been trained.

Operator errors vs Equipment Operator errors vs Equipment

Double exposure Double exposure

Hand held fluoro device ? Hand held fluoro device ?