>Radiation Protection in Paediatric Radiology Understanding Radiation Units L02 Radiation Protection in Paediatric Radiology Understanding Radiation Units L02

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 2 2 Educational Objectives At the end of the programme, the participants should become familiar with the following: Why is it important to measure radiation dose in children? How radiation dose can and should be expressed? Understand the radiation quantities and units used in diagnostic radiology.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 3 3 Answer True or False The same amount of radiation falling on the person at level of breast, head or gonads will have the same biological effects. Effective dose can be easily measured. Diagnostic reference levels are not applicable to paediatric radiology.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 4 Contents Dose descriptors outside the patient’s body. Dose descriptors for effects that have threshold (deterministic effects) Dose descriptors to estimate stochastic risks Diagnostic reference levels Dose descriptors and units for staff dose assessment

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 5 5 Introduction Several quantities and units are used in the field of diagnostic radiology to measure and describe radiation dose Some can be measured directly while others can only be mathematically estimated

>6 Two types of radiation effects Stochastic effects Where the severity of the result 6 Two types of radiation effects Stochastic effects Where the severity of the result is the same but the probability of occurrence increases with radiation dose, e.g., development of cancer There is no threshold for stochastic effects Examples: cancer, hereditary effects Deterministic effects Where the severity depends upon the radiation dose, e.g., skin burns The higher the dose, the greater the effect There is a threshold for deterministic effects Examples: skin burns, cataract Radiation Protection in Paediatric Radiology L02. Understanding radiation units

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 7 Hot Coffee – Energy contained in a sip Excess Temperature = 60º - 37 = 23º 1 sip = 3ml 3x 23 = 69 calories

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 8 Radiation Dose Lethal Dose= 4Gy LD 50/60 = 4 Gy For man of 70 kg Energy absorbed = 4 x 70 = 280 J = 280/418= 67 calories = 1 sip Energy content of a sip of coffee if derived in the form of X-rays can be lethal X-rays

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 9 9 Dose of Radiation Radiation energy absorbed by a body per unit mass.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 10 Dose Quantities and Radiation units - Dose quantities external to the patient’s body. - Dose quantities to estimate risks of skin injuries and effects that have threshold. - Dose quantities to estimate stochastic risks.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 11 Why so many quantities? Radiation dose is a complex topic 1000 Watt heater giving off heat (IR radiation) - unit is of power which is related with emission intensity Heat perceived by the person will vary with so many factors: distance, clothing, room temperature As can be seen with the example of heat, the energy transformation is a highly complicated issue This is the case with X-rays - radiation can’t be perceived

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 12 12 Basic Radiation Quantities Used to quantify a beam of X or γ-rays There are: Quantities to express total amount of radiation. Quantities to express radiation at a specific point

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 13 Exposure: X Exposure is a dosimetric quantity for measuring ionizing electromagnetic radiation (X-rays & Ɣ-rays), based on the ability of the radiation to produce ionization in air. Units: coulomb/kg (C/kg) or roentgen (R) 1 R = 0.000258 C/kg

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 14 KERMA KERMA (Kinetic Energy Released in a Material): Is the sum of the initial kinetic energies of all charged ionizing particles liberated by uncharged ionizing particles in a material of unit mass For medical imaging use, KERMA is usually expressed in air SI unit = joule per kilogram (J/kg) or gray (Gy) 1 J/kg = 1 Gy

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 15 Absorbed dose, D, is the mean energy imparted by ionizing radiation to matter per unit mass. SI unit = joule per kg (J/kg) or gray (Gy). In diagnostic radiology, KERMA and D are equal. Absorbed dose: D Harold Gray

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 16 Mean absorbed dose in a tissue or organ The mean absorbed dose in a tissue or organ DT is the energy deposited in the organ divided by the mass of that organ.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 17 17 Now things get a little more complicated !

>Radiation Dose Quantities Primary physical quantities are not used directly for dose limitation The Radiation Dose Quantities Primary physical quantities are not used directly for dose limitation The International Council on Radiation Protection (ICRP) has defined values for dose limits in occupational exposure Radiation Protection in Paediatric Radiology L02. Understanding radiation units

>Radiation Dose Quantities Equivalent  Dose:  Accounts for the type of radiation Different Radiation Dose Quantities Equivalent Dose: Accounts for the type of radiation Different radiation types have different level of biologic damage per unit absorbed dose Radiation Protection in Paediatric Radiology L02. Understanding radiation units

>Radiation Weighting Factors, wR (Source: ICRP 103)   Radiation Protection in Paediatric Radiology Radiation Weighting Factors, wR (Source: ICRP 103) Radiation Protection in Paediatric Radiology L02. Understanding radiation units

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 21 Equivalent Dose : HT,R The absorbed dose in an organ or tissue multiplied by the relevant radiation weighting factor : where DT,R is the average absorbed dose in the organ or tissue T, and wR is the radiation weighting factor for radiation R.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 22 22 Radiation Quantities and Units Equivalent dose (Unit = sievert, Sv ) Compares the biological effects for different types of radiation, X-rays, Ɣ-rays, electrons, neutrons, protons, α-particles etc. For X-rays, Ɣ-rays, electrons : absorbed dose and equivalent dose have the same value Gy = Sv. Rolph Sievert

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 23 Detriment Radiation exposure to different organs and tissues in the body results in different probabilities of harm and different levels of severity. The combination of probability and severity of harm is called “detriment”. Effective dose reflects the combined detriment from stochastic effects due to the equivalent doses in all the organs and tissues of the body.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 24 24 Effective Dose: ET Effective dose takes into account the organ specific radio-sensitivity to develop cancer and hereditary effects from radiation Unit = sievert, Sv

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 25 Effective Dose: ET A summation of the tissue equivalent doses, each multiplied by the appropriate tissue weighting factor: where HT is the equivalent dose in tissue T and wT is the tissue weighting factor for tissue T.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 26 26 The organs have different weighting factors, wT. These factors are published in ICRP 103 (2007) and have been changed over the years due to increased knowledge. Tissue Weighting Factors, wT

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 27 27 The weighting factors sum up to 1.0. They are relative and compares one organ with the other. They are the same for children and adults! Tissue Weighting Factors

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 28 Data is primarily taken from knowledge derived from studying the Japanese population exposed to atomic bombs in Hiroshima and Nagasaki On going research has changed the weighting factors from 1990 (ICRP 60) to 2007 (ICRP 103). 28 Tissue Weighting Factors

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 29 Multipliers of the equivalent dose to an organ or tissue to account for the different sensitivities to the induction of stochastic effects of radiation. Tissue Weighting Factors *ICRP 103 **Remainder Tissues (14 in total): Adrenals, Extrathoracic (ET) region, Gall bladder,Heart, Kidneys, Lymphatic nodes, Muscle, Oral mucosa, Pancreas, Prostate, Small intestine, Spleen, Thymus, Uterus/cervix..

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 30 30 Dose to lungs times their weighting factor; DL x wL + Dose (mean absorbed dose) to gastrointestinal tract times their weighting factor; DGI x wGI + ....(summation over organ after organ) = Effective dose Effective Dose (E) where T stands for tissue

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 31 31 Effective Dose (E) We can compare different paediatric imaging procedures through their different effective doses, E.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 32 32 Radiation Quantities and Units used in Diagnostic Radiology Incident air kerma Entrance surface air kerma Air kerma-area product Air kerma-length product Dosimetric quantities for CT Dosimetric quantities for interventional radiology

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 33 Incident Air Kerma Measured Free in Air on the central beam axis at the focal spot to surface distance. Only primary beam is considered, that is, no scatter contribution. Unit: joule/kg or gray (Gy)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 34 Entrance Surface Air Kerma (ESAK) ESAK measured on the surface of the patient or phantom where X-ray beam enters the patient or phantom. Includes a contribution from photons scattered back from deeper tissues, which is not included in free in air measurements.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 35 Entrance Surface Air Kerma (ESAK) If measurements are made at other distances than the true focus - to - skin distance, doses must be corrected by the inverse square law and backscatter factor incorporated into the calculation. References: Dosimetry in Diagnostic Radiology: An International code of practice, TRS 457, IAEA, 2007 Phys. Med. Biol. 43 (1998) 2237-2250.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 36 36 Kerma in X-ray field can be measured using calibrated: Ionization chamber Semiconductor dosimeter Thermoluminescent dosimeter (TLD) Dose Measurement

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 37 Kerma-Area Product: KAP The kerma - area product (KAP) is defined as the kerma in air in a plane perpendicular to the incident beam axis, integrated over the area of interest. This is the dose related quantity measured and displayed on all modern X-ray equipment excluding CT. KAP meter

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 38 Kerma-Area Product: KAP The KAP (Gy·cm2) is constant with distance since the cross section of the beam is a quadratic function which cancels the inverse quadratic dependence on dose . KAP remains constant along the beam axis as long as it is not measured close to the patient/phantom surface which introduces backscatter.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 39 39 KAP = K x Area the SI unit of KAP is the Gy·cm2 Kerma-Area Product: KAP

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 40 40 KAP is independent of distance from the X-ray source, as: Air Kerma decreases with the inverse square law. Area increase with the square distance KAP is usually measured at the level of the tube diaphragms Kerma-Area Product: KAP

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 41 This is a picture of a KAP meter which measures the kerma area product

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 42 42

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 43 In paediatric radiology KAP may be used for: Diagnostic reference levels (DRLs) By use of conversion factors, it can be converted to skin dose and/or effective dose 43 Kerma-Area Product

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 44 Dosimetric Quantities for CT Computed Tomography Dose Index (CTDI) - determined using scan protocol parameters. -useful for comparison of different scanners. Dose-Length Product (DLP) - measure of dose to patient - used to estimate effective dose

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 45 CT and Risk KERMA (in phantom) CTDI (dose in phantom per slice) Length of scan and pitch DLP Effective dose Risk

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 46 Measurement of Dosimetric Quantities in CT Pencil ionisation chamber with active length of 100 mm. Measurements free-in-air or in standard dosimetry phantom. Alternatives: TLD, solid state detectors. CTDIVOLshould be displayed on the console, reflecting the conditions of operation selected (IEC, 2003)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 47 Dose Indicators in Interventional Radiology For quality assurance purposes To estimate the probability of occurrence of stochastic effects use: Kerma-air product rate (KAP, PKA)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 48 Dose Indicators in Interventional Radiology For quantifying the threshold and severity of deterministic effects use: Maximum skin dose (MSD) Cumulative dose (CD) to Interventional Reference Point (IRP) In a complex procedure skin dose is highly variable

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 49 49 In some procedures, patient skin doses approach those used in radiotherapy fractions Maximum skin dose (MSD) or peak skin dose is the maximum dose received by a portion of the exposed skin. Interventional Procedures: Skin Dose Radiodermatitis in the right arm. 7 year-old patient. Photograph taken 4 months after radiofrequency ablation. Surce: ICRP 85

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 50 Cumulative Dose to Interventional Reference Point* IRP is located 15 cm from the isocentre towards the focal spot The air kerma accumulated at a specific point in space relative to the fluoroscopic gantry (IRP) during a procedure Cumulative dose does not include tissue backscatter and is measured in Gy. Cumulative dose is sometimes referred to as cumulative air kerma *IRP

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 51 Cumulative dose to Interventional Reference Point Cumulative dose to IRP is measured with a flat ion chamber or calculated by the system and displayed in the angiography room

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 52 MSD vs. Cumulative dose In some procedures, cumulative dose to IRP is well correlated with MSD Cumulative dose to IRP can be a good indicator of doses higher than the thresholds for skin injures A “trigger value” for cumulative dose can be adopted to alert interventionalists the threshold for skin erythema could be reached. A follow-up protocol can be adopted.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 53 53 Other related dose parameters Fluoroscopy time: Has a weak correlation with KAP But, in a quality assurance programme it can be adopted as a starting unit for comparison between operators, centres, procedures for the evaluation of protocol optimization, and to evaluate operator skill

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 54 54 Other related dose parameters Number of acquired images and number of series: Patient dose is a function of total acquired images But dose/image can have big variations There is an evidence of large variation in protocols adopted in different centres

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 55 Diagnostic reference levels (DRLs) ICRP, IAEA, EC: introduced the concept of diagnostic reference levels (DRLs) for patients DRLs are a form of investigation level, apply to an easily measured quantity at the surface of a simple standard phantom or a representative patient. An optimisation tool, not dose limits

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 56 Diagnostic Reference Levels (DRLs) DRLs calls for local investigation (often very simple) if constantly exceeded DRLs: Management of patient doses must be consistent with the required clinical imaging information

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 57 Quantities for Establishment of DRLs Incident air kerma and entrance-surface air kerma Incident air kerma rate and entrance-surface air kerma rate Air kerma–area product CT Dose index, CT Dose–length product

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 58 Quantities and Units for Staff Dose Assessment Personal dosimetry services typically provide monthly estimates of Hp(10) (mSv), the dose equivalent in soft tissue at 10 mm depth. This is in most of the cases used to estimate the effective dose. Sometimes, Hp(0.07) (mSv) is also reported: the dose equivalent in soft tissue at 0.07 mm depth) Personal dosememters (film, thermoluminescent...)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 59 Personal Dosimetry Methods Single dosimeter worn above the apron at neck level (recommended) or under the apron at waist level Two dosimeters worn (recommended in intrevational procedures) one above the apron at neck level another under the lead apron at waist level

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 60 Dose due to scatter radiation at a point occupied by the operator can be measured with a portable ionization chamber Dose Measurement

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 61 Summary Dosimetric quantities are useful to know the potential hazard from radiation and to determine radiation protection measures to be taken Physical quantities - Directly measurable Protection quantities - Defined for dose limitation purposes, but not directly measurable. Application specific quantities - Measurable in medical imaging. Diagnostic Refernce Levels

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 62 62 Answer True or False The same amount of radiation falling on the person at level of breast, head or gonads will have same biological effects. Effective dose can be easily measured. Diagnostic reference levels are not applicable to paediatric radiology.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 63 63 Answer True or False False -Different organs have different radio-sensitivity and tissue weighting factors as given by ICRP. False -It can be only calculated using different methods. False - DRLs apply for paediatric radiology, but these are age-specific.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 64 References INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS, Patient Dosimetry for X Rays Used in Medical Imaging, ICRU, Rep. 74, ICRU, Bethesda, MD (2006). INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Radiological Protection in Medicine, Publication 105, Elsevier, Oxford (2008) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Recommendations of the ICRP, Publication 103, Elsevier, Oxford (2008) EUROPEAN COMMISSION, Guidance on Diagnostic Reference Levels (DRLs) for Medical Exposure, Radiation Protection 109, Office for Official Publications of the European Communities, Luxembourg (1999) INTERNATIONAL ATOMIC ENERGY AGENCY, Dosimetry in Diagnostic Radiology: an International Code of Practice, Technical Report Series No 457, IAEA, Vienna (2007)

>Additional information Additional information

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 66 Quantities for radiation measurement Physical quantities - Directly measurable Protection quantities - Defined for dose limitation purposes, but not directly measurable Application specific quantities - Measurable in medical imaging

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 67 Radiation quantities and units Fundamental dosimetric quantities Protection quantities Equivalent dose Effective dose Application specific dosimetric quantities used in DR Incident air kerma Entrance surface air kerma Air kerma area product Air kerma length product Dosimetric quantities in CT and mammography

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 68 Physical Quantities

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 69 Physical quantities Fluence Exposure Kerma Absorbed dose

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 70 Fluence : f The fluence, f , is the quotient of dN by da, where dN is the number of particles incident on a sphere of cross section da, thus f = dN/da The unit of fluence is m-2

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 71 Exposure: X where dQ is the absolute value of the total charge of ions produced in air when all the electrons liberated in air of mass dm are completely stopped in air. The SI unit of exposure is the coulomb per kilogram (C/kg) The special unit of exposure is the röntgen (R). 1R = 2.58 x 10-4 C kg-1

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 72 KERMA The KERMA (Kinetic Energy Released in a MAterial) where dEtrans is the sum of the initial kinetic energies of all charged ionizing particles liberated by uncharged ionizing particles in a material of mass dm The SI unit of kerma is the joule per kilogram (J/kg), termed gray (Gy). .

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 73 Exposure and KERMA Exposure, X, in units of C kg-1, is related to air kerma as follows: where W is the average energy spent by an electron to produce an ion pair, g is the fraction of secondary charged particles that is lost to bremsstrahlung radiation production and e is the electronic charge

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 74 The fundamental dosimetric quantity absorbed dose, D, is defined as: where is the mean energy imparted by ionizing radiation to matter in a volume element and dm is the mass of matter in the volume element. The SI unit of absorbed dose is the joule per kilogram (J/kg), termed the gray (Gy) In diagnostic radiology, KERMA and D are equal Absorbed Dose: D

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 75 Exposure and Absorbed Dose or KERMA Exposure can be linked to air dose or kerma by suitable conversion coefficients. For example, 100 kV X-rays that produce an exposure of 1 R at a point will also give an air kerma of about 8.7 mGy and a tissue kerma of about 9.5 mGy at that point.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 76 Application Specific Quantities

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 77

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 78 Application Specific Quantities Schematic diagram showing some dosimetric and geometric quantities

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 79 Entrance Surface Air Kerma (ESAK) where Y(kVp, FFD) is tube output for actual kVp used during examination, mAs is actual tube current-time product used during examination and FFD is focus-to-film distance. BSF is the backscatter factor that depends on kVp and total filtration of X-rays

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 80 Backscatter Factors (Water)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 81 81 If the KAP is calculated by the system, you must know if the user added filtration you use is included or not ! Kerma-Area Product: KAP

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 82 Kerma-Area Product: KAP It is always necessary to calibrate and to check the transmission chamber for the X-ray installation in use In some European countries, it is compulsory that new equipment is equipped with an integrated ionization transmission chamber or with automatic calculation methods

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 83 Dosimetric Quantities for CT Computed Tomography Dose Index (CTDI) CT air kerma index Dose-Length Product (DLP) Air kerma-length product

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 84 ICRU 74 / IAEA TRS 457 CT air kerma index Free-in-air (Ck) In phantom (Ck,PMMA) Air kerma length product (PKA)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 85 Dosimetric Quantities for CT Principal dosimetric quantity in CT is CT air kerma index: where K(z) is air kerma along a line parallel to the axis of rotation of the scanner over a length of 100 mm. N = Number of detectors in multi-slice CT T = Individual detector dimension along z-dimension The product NT defines the nominal scan beam width/collimation for a given protocol.

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 86 Dosimetric Quantities for CT Weighted CT air kerma index, CW, combines values of CPMMA,100 measured at the centre and periphery of a standard CT dosimetry phantoms

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 87 Dosimetric Quantities for CT Pitch (IEC, 2003): T= Single detector dimension along z-axis in mm. N=Number of detectors used in a given scan protocol (N>1 for MDCT), N x T is total detector acquisition width or collimation I=table travel per rotation Radiographic, 2002, 22:949-62

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 88 Volume CTDI describes the average dose over the total volume scanned in sequential or helical sequence, taking into account gaps and overlaps of dose profiles (IEC, 2003): Average dose over x, y and z direction Protocol-specific information Dosimetric Quantities for CT

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 89 Dosimetric Quantities for CT Kerma-length product (PKL): where L is scan length is limited by outer margins of the exposed scan range (irrespective to pitch) PKL for different sequences are additive if refer to the same type of phantom (head/body)

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 90 Maximum Skin Dose (MSD) Measurement/evaluation of MSD Point or area detectors Cumulative dose at IRP (interventional radiology point) Calculation from technical data Off line methods Area detectors: TLD array, slow films, radiochromic films From KAP and Cumulative dose measurement

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 91 Method for MSD Evaluation: Radiochromic Large Area Detector Example: Radiochromic films type Gafchromic XR R 14”x17” useful dose range: 0.1-15 Gy minimal photon energy dependence (60 - 120 keV) acquisition with a flatbed scanner:b/w image, 12-16 bit/pixel or, measure of OD measurement with a reflection densitometer

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 92 Benefits of Radiochromic Films The radiochromic film: displays the maximum dose and its location shows how the total dose is distributed provides a quantitative record for patient files provides physician with guidance to enable safe planning of future fluoroscopically guided procedures improves fluoroscopic technique and patient safety possible rapid semi-quantitative evaluation Example of an exposed radiochromic film in a cardiac interventional procedure

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 93 Rapid Semi-Quantitative Evaluation: Example For each batch number (lot #) of gafchromic film a Comparison Tablet is provided In the reported example we easily can recognise that the darkness area of the film, corresponding to the skin area that has received the maximum local dose, has an Optical Density that correspond at about 4 Gy

>Radiation Protection in Paediatric Radiology         Radiation Protection in Paediatric Radiology L02. Understanding radiation units 94 94 DRLs for Complex Procedures 3rd level “Patient risk” 2nd level “Clinical protocol” 1st level “Equipment performance” Reference levels (indicative of the state of the practice): to help operators to conduct optimized procedures with reference to patient exposure For complex procedures reference levels should include: more parameters and, must take into account the complexity of the procedures. (European Dimond Consortium recommendations)